marekf 2660c564a6 <Feat>: Expose Genesis preadjustment results in Results and improve result mapping
Cause:

- Preadjustment values were not available for persistent Results mapping.
- Older Genesis item IDs could resolve to different result variables.
- Meter-specific variables were difficult to find.

Solution:

1. Added preadjustment result storage
   - Store common execution metadata and available phase-specific values per slot and test.
   - Expose temperature comparison and Completion values in Results.
   - Preserve existing measurement execution, timing, parallelism and STOP handling.

2. Improved result selection
   - Add Water meter type under the existing Category tab.
   - Group available items into Common, Genesis and iPerl.

3. Corrected saved mappings
   - Migrate recognized legacy Genesis mappings.
   - Store stable item identifiers while preserving existing numeric IDs.
   - Mark ambiguous mappings as Review mapping.
   - Preserve the one-based Genesis repetition number.
2026-09-14 22:48:10 +02:00
marekf 7e2b4ed908 Merge branch 'feature/task/UnionTown-Cordonel-completion-at-the-customer-last-task-6' into feature/task/UnionTown-Cordonel-completion-at-the-customer-milestone 2026-09-14 19:31:10 +02:00
marekf c1960f2150 <Feat>: Add per-slot preadjustment file logs and diagnostic headers, increase revision to 3.9.3146.110
Cause:

- Preadjustment diagnostics required persistent log files for each slot.
- Phase boundaries and file storage outcomes were not clearly visible.

Solution:

1. Added per-slot file logging
   - Store logs under C:\Tbf\Logs\Preadjustment.
   - Create a separate directory for each batch and a separate file for each slot.
   - Capture phase diagnostics, Completion SQL messages and STOP requests.
   - Support normal execution and simulation independently of the diagnostic window.

2. Added phase formatting
   - Add batch and phase headers with slot and start time.
   - Include phase result and duration in file footers.
   - Display headers in the corresponding slot diagnostic window without duplicate file entries.

3. Added file storage diagnostics
   - Display the full log path and creation/write status per slot.
   - Confirm successful writes after closing the file.
   - Report write failures and incomplete logs.

4. Added background file writing
   - Buffer and write log entries without blocking serial reception.
   - Preserve parallel slot execution and existing calibration behaviour.

5. Increased revision
   - Updated revision from 3.9.3146.109 to 3.9.3146.110.
2026-09-14 19:00:39 +02:00
marekf 4124443f36 <Feat>: Add Completion and Q3 calibration database writes through GCI, increase revision to 3.9.3146.109
Cause:

- Preadjustment Completion and TBF Q3 calibration results required configurable database persistence.
- Both write operations required diagnostics and simulation support.

Solution:

1. Added independent database writer configurations
   - Added PreadjustmentCompletionWriter and Q3CalibrationWriter to gci_config.
   - Support SQL commands, stored procedures and parameter mappings.
   - Added SQL table creation and verification scripts.

2. Added Preadjustment Completion storage
   - Store PCB ID, meter size, amplitude settings, temperature offsets, zero-flow offsets, timestamp and overall result.
   - Use captured Completion readout values.
   - Preserve the original HTTP reporting.
   - Report SQL connection and write status in the corresponding slot diagnostics.

3. Added Q3 calibration storage
   - Forward TBF results through GCI Bridge to the GCI database API.
   - Store signed deviations and written calibration factors for all three paths.
   - Write to SQL after successful calibration register writes and StoreCalibration.
   - Include PCB ID and UTC timestamp.

4. Added simulation support
   - Generate sample Completion and Q3 records for simulated slots.
   - Perform real SQL writes when the corresponding writer is enabled.
   - Log connection progress, write outcomes and generated record IDs.

5. Separated calibration and database outcomes
   - Preserve meter operation results when database storage fails.
   - Expose database write status independently.

6. Increased revision
   - Updated revision from 3.9.3146.108 to 3.9.3146.109.

Validation:

- Confirmed successful SQL storage for simulated Completion.
- Confirmed successful Q3 calibration result storage.

ToDo:

- Use the GCI Q3 storage API as the foundation for future migration of Q3 calibration from the direct TBF communication branch to GCI.
2026-09-14 17:06:22 +02:00
marekf 6bab62a062 <Fix>: Integrate GCI and direct streaming workflows, restore parallel preadjustment execution and update revision to 3.9.3146.108
Cause:

- GCI streaming port ownership interfered with regular TBF measurements using direct streaming access.
- Streaming handoff recreated both connections and required unnecessary login attempts between phases.
- Sequential port acquisition delayed execution, while unavailable meters could prevent healthy slots from continuing.
- Repeated initialization retained exclusions from the previous attempt.

Solution:

1. Integrated two communication approaches
   - Enable coexistence of the GCI Bridge workflow, which uses request and streaming ports, and the TBF workflow, which accesses the streaming port directly.
   - Transfer streaming port ownership explicitly between both workflows.
   - Preserve the GCI request connection and authentication during streaming handoff.

2. Updated ExternalProjects - GenesisCore/GenesisMeter
   - Added ConnectStreamingPort and DisconnectStreamingPort.
   - Separate streaming port lifecycle from request connection lifecycle.
   - Allow streaming port ownership to move between both technological branches without reconnecting the request port or repeating login.
   - Preserve process configuration and release streaming resources on completion, cancellation or acquisition failure.

3. Updated ExternalProjects - CordonelPreadjustmentUi
   - Adapt ZeroFlowGenesisMeter data collection to streaming port handoff.
   - Support repeated StartRecordData calls without duplicate subscriptions.
   - Prevent duplicate record processing and repeated amplitude scaling.
   - Coordinate preadjustment data subscription startup and shutdown with streaming ownership.

4. Restored parallel slot handling
   - Acquire, check and release ports concurrently across slots.
   - Isolate unavailable or failed meters without excluding healthy slots.
   - Drain pending operations before cleanup on STOP.

5. Corrected repeated execution and results
   - Reset exclusions and result state on explicit preadjustment initialization.
   - Evaluate results separately for each slot.
   - Prevent empty phases from being reported as successful measurements.
   - Return the streaming port for subsequent non-GCI measurements.

6. Updated revision
   - Set revision to 3.9.3146.108.

Related dependency changes:

- This commit requires the corresponding changes in:
  - ExternalProjects - GenesisCore/GenesisMeter.
  - ExternalProjects - CordonelPreadjustmentUi.
- Build and deploy both updated assemblies together with the matching TBF/GCI changes.

Architecture milestone:

- Established functional integration of two communication architectures: GCI Bridge access to both ports and direct TBF access to the streaming port.
- This version, together with the corresponding ExternalProjects changes, provides the baseline for their future unification.

ToDo:

- Use this baseline to unify request and streaming communication behind the GCI interface.
- Migrate remaining direct streaming access to GCI while preserving parallel slot execution, cancellation and diagnostic logging.
- Centralize ownership and lifecycle management of both ports in GCI.
2026-09-12 00:23:37 +02:00
marekf c426df990c <Feat>: Add per-slot temperature calibration results, diagnostic hold and settling delay, increase revision to 3.9.3146.107
Cause:

- Temperature calibration results lacked measured meter temperatures and comparison details.
- Diagnostics closed immediately after a phase, preventing inspection of failures.
- Push Temperature remained visible when temperature was supplied directly from T1/T2.
- Automatic phase transitions provided insufficient settling time before temperature calibration, causing temperature checks to fail.

Solution:

1. Added structured temperature calibration results
   - Return reference temperature, meter temperature, absolute difference, effective limit and sample count per slot.
   - Distinguish temperature check status from overall calibration completion.
   - Preserve existing API signatures and add typed result methods.

2. Improved temperature diagnostics
   - Log sample count, invalid values, minimum, maximum and mean for each path.
   - Add bounded TEMP-RAW capture with raw temperature, scaling factor, decoded temperature, timestamps and frame content.
   - Attach diagnostic capture only during the temperature check and detach it afterwards, including on cancellation.
   - Preserve the original streaming handler used during Amplitude Test.
   - Keep the original comparison algorithm and temperature deviation limit.
   - Remove configurable deviation from the GCI Bridge component dialog.

3. Added optional hold after preadjustment
   - Add the Hold after preadjustment procedure parameter.
   - Keep completed results and diagnostics visible until Continue is pressed.
   - Process STOP during hold and exclude hold time from measurement duration.

4. Updated phase display
   - Hide Push Temperature in sensor mode unless that action is explicitly running.
   - Keep failed slots and their exclusion status visible during hold.

5. Corrected settling time
   - Remove initialization overrides that forced OffsetTestSettlingTime to zero.
   - Set OffsetTestSettlingTime to 20 seconds.
   - Apply settling before temperature calibration and during Offset Test.
   - Preserve STOP handling during the wait.
   - The existing inclusive countdown produces approximately 21 seconds of waiting.

6. Aligned project versions
   - Update TBF revision to 3.9.3146.107.
   - The corresponding Laatzen PreadjustmentUI project state matches this TBF revision exactly.
   - Build and deploy these matching project versions together.

Validation:

- Compared raw temperature frames from GCI GUI and GenesisCommunication.
- Confirmed that decoded Meter temperatures matched the raw frame values.
- Confirmed that adding a pause after Amplitude Test allowed temperature calibration to pass.
2026-09-11 19:05:17 +02:00
marekf 20614fb9e6 <Fix>: Add real-time slot progress logging to Offset Test, increase revision to 3.9.3146.106
Cause:

- Settling and data collection updated only StatusLabel, leaving slot diagnostic memos without progress messages.
- One error message had no slot assigned and was ignored by slot diagnostics.

Solution:

1. Added slot-specific progress logging
   - Log settling countdown every second.
   - Log collected record count every collection cycle.
   - Log completion of data collection.

2. Corrected error routing
   - Include slot and PCB ID in the bad-data warning.

3. Preserved test behaviour
   - Keep existing timing, calculations and validation unchanged.

4. Increased revision
   - Updated revision to 3.9.3146.106.
2026-09-11 03:49:31 +02:00
marekf 302431b64c <Fix>: Add skipped preadjustment phase state and improve diagnostic colors, increase revision to 3.9.3146.105
Cause:

- Optional preadjustment phases remained in the Pending state after a later phase had already started.
- Running and skipped phases required clearer visual distinction.

Solution:

1. Improved preadjustment phase states
   - Mark preceding unexecuted phases as Skipped when a later preadjustment phase starts.
   - Allow a skipped phase to transition normally to Running if it is executed later.

2. Improved diagnostic colors
   - Display running phases in blue.
   - Display skipped phases in orange.
   - Preserve green for passed, red for failed and grey for pending phases.

3. Increased revision
   - Updated revision to 3.9.3146.105.
2026-09-11 03:12:31 +02:00
marekf db34c579d0 <Fix>: Correct GCI streaming port ownership for preadjustment, increase revision to 3.9.3146.104
Cause:

- GenesisRegReader configured the GCI streaming port as NA.
- GCI Bridge therefore received no calibration streaming data during the Amplitude Test.
- The request port remained functional, so register writes succeeded while TOF data collection failed.

Solution:

1. Corrected streaming port ownership
   - Pass the configured Genesis opto COM port to GCI Bridge as the UART streaming port.
   - Stop GenesisRegReader streaming before handing the port over to GCI Bridge.
   - Validate request and streaming ports during preadjustment detection.
   - Release the GCI connection after preadjustment so the port can be reused by subsequent tests.

2. Increased revision
   - Updated revision to 3.9.3146.104.
2026-09-11 02:49:09 +02:00
marekf 83afddc918 <Feat>: Add real-time Genesis preadjustment diagnostics and configurable per-slot simulation, increase revision to 3.9.3146.103
Cause:

* Genesis preadjustment activities lacked a dedicated real-time diagnostic view.
* It was difficult to monitor slot workers, requests, phase progress and slot-specific logs during execution.
* GCI Bridge simulation did not provide configurable timing and results for individual preadjustment activities.
* Failed slots could incorrectly continue in subsequent preadjustment phases.
* Live log refresh prevented reliable scrolling and copying while messages were arriving.

Solution:

1. Added preadjustment diagnostic logging

   * Added detailed logging for GenesisCommunication preadjustment execution.
   * Added GCI Bridge worker, queue, request and operation diagnostics.
   * Routed diagnostic messages through AppDiagnostic.

2. Added a dedicated preadjustment diagnostics form

   * Opens automatically while a preadjustment activity is running.
   * Closes automatically when the activity finishes.
   * Uses a stable position within the TBF results area.
   * Displays GCI slot and worker status in real time.

3. Added dynamic slot diagnostics

   * Creates a tab for every selected slot.
   * Displays all eight preadjustment phases.
   * Shows Pending, Running, Passed and Failed states using distinct colors.
   * Preserves phase history between independently executed activities.
   * Keeps failed and excluded slots visible.

4. Added slot-specific live logs

   * Routes progress and simulation messages to the corresponding slot memo.
   * Appends only new messages instead of replacing the complete text.
   * Preserves the operator's scroll position while new messages arrive.
   * Removed duplicate timestamps.

5. Added configurable preadjustment simulation

   * Added configurable simulation duration, Success and ErrorMessage parameters.
   * Added per-operation simulation profiles to GCI Bridge.
   * Applies procedure-specific simulation values dynamically before executing a GCI operation.
   * Rejects invalid simulation-mode combinations.

6. Added parallel per-slot simulation workers

   * Executes simulated preadjustment operations independently for every selected slot.
   * Uses a dedicated worker and queue state for each slot.
   * Waits for all participating operations using Task.WhenAll.
   * Aggregates individual slot results into the batch result.

7. Added failed-slot exclusion

   * Stores preadjustment results for the current batch.
   * Excludes failed slots from subsequent preadjustment activities.
   * Keeps excluded slots visible and marked as failed.
   * Limits exclusion to preadjustment activities.

8. Increased revision

   * Updated revision to 3.9.3146.103.

Validation:

* Confirmed real-time phase and slot diagnostics.
* Confirmed parallel simulation using dedicated slot workers.
* Confirmed configurable simulation timing and results.
* Confirmed live slot logs remain scrollable and copyable.
2026-09-11 02:36:17 +02:00
marekf cf77faf972 kick 2 - prva dobra s dialogom testov 2026-09-11 00:33:16 +02:00
marekf 45a6d59b5e kick1 - priebezne stabilne - pred velkym zadanim 2026-09-10 21:39:00 +02:00
marekf 73c8fecdf2 <Update>: Version to 3.9.3146.102 in AssemblyInfo files 2026-09-10 17:52:22 +02:00
marekf 8c77df207a <Feat>: Add universal NLog diagnostics and improve AppDiagnostic UI
Cause:

- Live diagnostics required logging from multiple applications without blocking TBF.
- Shared nlog.config files caused configuration conflicts.
- The original viewer lacked configurable colors and practical log controls.
- Inconsistent diagnostic deployment caused log transfer failures and locked shared DLLs during builds.

Solution:

1. Added universal diagnostic transport
   - Added AppDiagnostic.Logging with an NLog target and named-pipe communication.
   - Added background buffering and automatic reconnection.
   - Registered the diagnostic connector programmatically.

2. Separated diagnostic window lifecycle
   - Run AppDiagnostic as an independent process.
   - Activate the existing window on repeated launches.
   - Keep TBF running when the diagnostic window closes.

3. Corrected diagnostic deployment
   - Copy the complete diagnostic output into a dedicated AppDiagnostic subdirectory during build.
   - Overwrite existing deployment files to prevent inconsistent versions.
   - Validate and launch the executable from that subdirectory with an explicit working directory.

4. Improved the log viewer
   - Added configurable substring colors with a palette and live preview.
   - Added text, level and application filters.
   - Added line wrapping, clipboard actions and log export.
   - Added Pause/Resume, Refresh, Clear, auto-scroll and Always on top.
   - Added connection and log counters.
   - Persist color rules between sessions.

5. Removed legacy dependencies
   - Replaced LiveLogCache with structured diagnostic storage.
   - Removed AppDiagnostic dependencies on log4net and SharedComponents.

Validation:

- Confirmed successful log transfer when launching AppDiagnostic from TBF.
2026-09-10 17:45:29 +02:00
marekf 1c6f9af2bb <Feat>: Add standalone Genesis preadjustment activities, fix configuration control and increase revision to 3.9.3146.101
Cause:

- GenesisCommunication did not expose GCI preadjustment phases as selectable test activities.
- Meter size and temperature inputs required by preadjustment could not be configured.
- The configuration control incorrectly expected the iPerlCommunication configuration type.
- Preadjustment results were not propagated to TBF test and meter results.

Solution:

1. Added standalone preadjustment activities
   - Added Initialization, Detect, Preparation, Amplitude Test, Temperature Calibration, Push Temperature, Offset Test and Completion.
   - Each phase can be selected and executed independently.

2. Added preadjustment parameters
   - Added configurable meter size and temperature.
   - Added validation and supported meter-size selection.

3. Added batch preadjustment execution
   - Resolves enabled Genesis register readers and their slots.
   - Verifies that all readers use the same GCI Bridge instance.
   - Dispatches the selected activity through the corresponding GCI Bridge API.

4. Added result processing
   - Stores execution time and completion status.
   - Updates test and meter results according to the GCI response.

5. Fixed the configuration control
   - Corrected TestMethodCfgCtrl to accept the GenesisCommunication TestMethodCfg type.

6. Increased revision
   - Updated revision to 3.9.3146.101.
2026-09-10 14:08:00 +02:00
marekf 0cfde7f06b <Update>: Version to 3.9.3146.0 in AssemblyInfo files 2026-09-09 17:31:27 +02:00
marekf d3c361b544 <Merge-fix>Solve merge conflicts - multichange: PassedFromTests() renamed to PassedFromAllTests() 2026-09-09 17:28:19 +02:00
marekf b35b7d92f9 <Merge-fix> Solve merge conflicts 2026-09-09 17:27:35 +02:00
marekf eb4f7ad971 <Fix>: Prevent component configuration crashes, add a copyable error dialog, increase revision to 3.9.3145.109
Cause:

* A configuration control can reject an incompatible configuration and return null.
* ComponentsManagerDlg accessed the rejected configuration without validation, causing a NullReferenceException and closing TBF.
* The standard MessageBox was difficult to read and did not provide convenient access to technical details.

Solution:

1. Added configuration validation

   * Validates the configuration returned by the component control.
   * Logs the rejected control and configuration types.
   * Cancels the add operation without terminating TBF.

2. Added a reusable error details dialog

   * Displays a clear error summary and structured technical details.
   * Allows the diagnostic text to be selected and copied.
   * Added Copy and Close actions and registered the new form in TBF.csproj.

3. Increased revision

   * Updated revision to 3.9.3145.109.
2026-09-09 10:09:49 +02:00
marekf 8677e49680 <Fix>: Correct GciBridge component class name generation, increase revision to 3.9.3145.108
Cause:

- GciBridge Factory used an incorrect namespace substring offset when generating the component class name.
- The generated value was `geComponents.GciBridge` instead of `BridgeComponents.GciBridge`.
- The invalid class name could prevent correct component registration and lookup.

Solution:

1. Corrected class name generation
   - Changed the namespace substring offset from 12 to 8.
   - GciBridge Factory now returns the correct `BridgeComponents.GciBridge` class name.

2. Restored component identification
   - GciBridge can now be registered and resolved using the expected TBF component class name.

3. Increased revision
   - Updated revision to 3.9.3145.108.
2026-09-08 22:00:31 +02:00
marekf 29faef86e3 <Fix>: Prevent GciBridge startup failure when storage components are unavailable, increase revision to 3.9.3145.107
Cause:

- GciBridge threw an exception when the configured UniDataStorageReader or UniDataStorageWriter was not available during component initialization.
- The exception prevented TBF from starting and made it impossible to correct the component configuration through the UI.

Solution:

1. Improved missing component handling
   - Replaced fatal exceptions for missing storage components with warning logs.
   - Allowed GciBridge and TBF to continue initialization.

2. Preserved existing behavior
   - Available UniDataStorageReader and UniDataStorageWriter components are still linked as before.
   - Existing GciBridge configurations remain compatible.

3. Increased revision
   - Updated revision to 3.9.3145.107.
2026-09-08 21:59:29 +02:00
marekf 6ef42fcc21 <Feat>: Add optional TestID filtering for repeating XML results, increase revision to 3.9.3145.106
Cause:

- Repeating XML output always included every published meter test.
- The existing TestID/Prüfpunkt configuration was not used to filter generated XML test records.
- A configured TestID could be evaluated in an incorrect repeating test context.

Solution:

1. Added optional test filtering
   - Non-empty TestID values define the tests included in repeating XML output.
   - Multiple configured TestID values act as a whitelist.

2. Preserved existing behavior
   - When no TestID is configured, all published tests are exported as before.
   - Existing configurations remain compatible without migration.

3. Prevented cross-test evaluation
   - Test-dependent items are evaluated only for their matching TestID.
   - Values from one test are no longer reused in another test record.

4. Improved source formatting
   - Reformatted ResultsWriter.cs for easier reading and searching.

5. Increased revision
   - Updated revision to 3.9.3145.106.
2026-09-08 14:11:09 +02:00
marekf d2d41df155 <Fix>: Restore parent UniDataStorageWriter in Previous results editor, increase revision to 3.9.3145.105
Cause:

- Previous results opens the ResultsWriter configuration through GetControl(null).
- The component entity list is therefore unavailable.
- The DropDownList could not display the saved ParentName because it was not present in its Items collection.

Solution:

1. Restored the saved parent writer
   - Adds the configured ParentName to the list when it is not already available.
   - Selects the restored UniDataStorageWriter in parentComboBox.

2. Preserved standard configuration behavior
   - All available UDSW components are still loaded when component entities are provided.
   - Previous results can now display and use the previously configured parent writer.

3. Improved source formatting
   - Removed inappropriate line wrapping from ResultsWriterCfgCtrl.cs.

4. Increased revision
   - Updated revision to 3.9.3145.105.
2026-09-08 13:20:55 +02:00
marekf 6d2477be4a <Feat>: Enable TestID column in ResultsWriter configuration, increase revision to 3.9.3145.104
Cause:

- ResultsWriter explicitly suppressed the existing TestID column.
- Users could not configure the test identifier for test-dependent result variables.
- In the German localization, the corresponding Prüfpunkt column was therefore hidden.

Solution:

1. Enabled TestID configuration
   - Set SupressTestIDColumn to false.
   - The TestID/Prüfpunkt column is now displayed after Merge.

2. Preserved existing configuration handling
   - TestID values continue to use the existing WMeterRsltItemSpec serialization.
   - No configuration format change was required.

3. Increased revision
   - Updated revision to 3.9.3145.104.
2026-09-08 12:54:53 +02:00
marekf 57152ae749 <Feat>: Add XML-based payload archive filename configuration, increase revision to 3.9.3145.103
Cause:

- Payload archive filenames required values from the generated XML payload.
- Date/time values needed filename-safe formatting.
- XML result values such as Passed and Failed needed configurable conversion to P and F.

Solution:

1. Added XML filename tokens
   - Supports any unique non-repeating XML value using {Xml:/path}.
   - Resolves filename values directly from the generated payload.

2. Added user-friendly token configuration
   - Clicking inside a token opens the XML value chooser.
   - Added Time format and Mask transformations.

3. Improved filename generation
   - Validates XML paths and rejects ambiguous repeating values.
   - Sanitizes resolved values for safe filesystem usage.
   - Passes the per-meter payload context from ResultsWriter.

4. Integrated the new dialog
   - Added PayloadFileNameXmlTokenDlg and its Designer file.
   - Updated the TBF project configuration.

5. Increased revision
   - Updated revision to 3.9.3145.103.
2026-09-08 12:52:40 +02:00
marekf 41408b30da <Feat>: Improve Previous Results layout and resizable result columns, increase revision to 3.9.3145.103
Cause:

- Previous Results used fixed column widths, making long procedure and purchase order values difficult to read.
- Result columns could not be resized interactively.
- Window resizing caused unnecessary UI redraws and visible lag.
- The existing result layout lacked clear column headers and visual separation.

Solution:

1. Added resizable result columns
   - Added draggable column separators.
   - Keeps column widths synchronized across all displayed result rows.
   - Added optimized resize preview to reduce UI lag.

2. Improved responsive layout
   - Enlarged the default Previous Results window.
   - Keeps Purchase Order, OK, NOK and Actions aligned to the right.
   - Uses Procedure as the flexible column when resizing the window.

3. Improved result presentation
   - Added aligned column headers.
   - Added modern column colors and styled action buttons.
   - Reduced spacing between result rows and improved separators.

4. Increased revision
   - Updated revision to 3.9.3145.103.
2026-09-04 08:25:27 +02:00
marekf 136be96e1b <Feat>: Add customer XML structure mapping support in ResultsWriter, increase revision to 3.9.3145.102
Cause:

- The customer XML format requires batch, DUT, group and test-level metadata.
- ResultsWriter did not provide variables for test position, repetition index, batch duration, DUT test duration or aggregated test statuses.
- Missing WaterMeterData could cause a NullReferenceException while evaluating WM Type.

Solution:

1. Added XML mapping variables
   - Added TestPositionInBatch and TestRepetitionNr.
   - Added BatchTime and TestsTotalTime.
   - Added PassedFromAllTests and PassedFromRepeatedTests.

2. Added hierarchical result evaluation
   - Evaluates the overall PANEL/DUT status from all tests.
   - Evaluates GROUP status from repeated instances of the selected test.

3. Improved meter data handling
   - Made WM Type evaluation null-safe when WaterMeterData is unavailable.

4. Increased revision
   - Updated revision to 3.9.3145.102.
2026-09-04 08:20:50 +02:00
marekf d3c8813012 <Fix>: Correct test-dependent result evaluation in ResultsWriter, increase revision to 3.9.3145.101
Cause:

- ResultsWriter evaluated selected test-dependent items without the actual TestID.
- This caused values such as Test passed(), timestamps, flow, pressure, temperature, conductivity and error data to be empty or incorrect.

Solution:

1. Fixed test-aware result evaluation
   - Uses published regular meter test results.
   - Passes mtr.Name() as TestID to item.Print(wm, testId).

2. Restored correct test result mapping
   - Test-dependent values are now resolved from the correct TestRslt / MeterTestRslt.

3. Increased revision
   - Updated revision to 3.9.3145.101.
2026-09-03 10:00:40 +02:00
marekf 054075a341 <Feat>: Add XML result generation with file and MSSQL output support, increase revision to 3.9.3145.100
Cause:

- ResultsWriter required a generic way to generate customer-specific XML result data from TBF measurement results.
- The customer reference XML contains example runtime values and repeated result structures, so it cannot be used directly as the generated output.
- TBF result variables must be explicitly mapped to destinations in the customer XML structure.
- The generated XML result data must support two output targets:
  - direct creation of an XML file,
  - delivery of the XML payload to a Microsoft SQL stored procedure.
- Preview generation must allow the XML structure and configured mappings to be verified without executing the production database write.
- Increase revision to 3.9.3145.100.

Solution:

1. Added XML reference analysis
   - Creates a clean base XML structure.
   - Extracts the repeating result prototype.
   - Prevents sample runtime values from the customer reference XML from leaking into generated results.
2. Added configurable TBF-to-XML result mapping
   - Allows explicit mapping of TBF result variables to customer XML destinations.
   - Supports one-time and repeating XML destinations.
   - Keeps the mapping independent of the semantic meaning of customer XML attribute names.
3. Added XML destination viewer and configurator
   - Shows the current mapping.
   - Highlights repeating destinations.
   - Identifies already used one-time destinations.
4. Added runtime XML result generation
   - Uses the configured TBF-to-XML mappings.
   - Uses measurement procedure result data.
   - Builds the output from the clean base XML and repeating XML prototype.
   - Generates repeated result records according to the executed measurement procedure.
5. Added simulation-based Preview request
   - Generates a complete XML payload using simulation values.
   - Allows XML structure and mapping verification before production execution.
   - Does not execute the production stored procedure.
6. Added direct XML file output support to UniDataStorageWriter
   - Supports File / .xml as a physical output target.
   - Creates the generated XML result file in the configured output directory.
7. Added Microsoft SQL stored-procedure XML output
   - Supports Microsoft SQL / StoredProcedure as a physical output target.
   - Passes the generated XML payload through the configured stored procedure parameter.
8. Added optional XML payload archiving
   - Allows generated XML payloads to be stored in the configured Payload archive.
   - Can be used together with the Microsoft SQL stored-procedure output.
9. Kept ResultsWriter independent of the physical output target
   - ResultsWriter generates the result payload.
   - UniDataStorageWriter decides how and where the payload is physically written.
   - The same ResultsWriter XML generation mechanism is therefore used for both XML file and MSSQL outputs.
10. Increased revision
   - Updated revision to 3.9.3145.100.
2026-09-02 12:05:10 +02:00
marekf a0d53f9ec9 <Fix>: RefPulses carry-over by restoring testInProgress behavior in StandingStart, increase revision to 3.9.3145
Cause:
1. Background: Restore/ensure testInProgress behavior before the DataEntry sequence and explicitly prevent to reset the pulse.
2. Increase revision to 3.9.3145

Solution:
1. Deleted .AddOperation(testInProgress) calling
2. Increased revision to 3.9.3145
2026-08-25 13:49:49 +02:00
marekf 1e45f48c75 <Update>: Version to 3.9.3144.0 in AssemblyInfo files 2026-08-25 08:23:27 +02:00
marekf d833eca584 <Merge_Fix>: System.Data.SQLite, System.Data.SQLite.Core, Stub.System.Data.SQLite.Core.NetFramework 2026-08-24 15:36:50 +02:00
marekf c966a7034c <Merge-Fix>: Solve merge conflicts 2026-08-24 15:04:11 +02:00
marekf f051d51ae0 <Fix>: Incorrect cycle resume behavior after Stop and Start, increase revision to 3.9.3134.114
Cause:
1. When the Cycle was stopped during the second or a later test and then resumed using the Start Test button, the resume confirmation was handled incorrectly. After selecting to continue, the sequence restarted from the first test instead of continuing from the interrupted test.
2. Increase revision to 3.9.3134.114

Solution:
1. Corrected the Cycle resume logic so that, after confirming continuation, the sequence resumes from the interrupted test instead of restarting from the first test.
2. Increased revision to 3.9.3134.114
2026-08-24 14:25:19 +02:00
marekf 267159e0a9 <Fix>: RefPulses carry-over by restoring testInProgress behavior in StandingStart, increase revision to 3.9.3134.113
Cause:
1. Background: there was a suspicion that the Bukoven board left RefPulses active across operations; this was investigated and ruled out. The pulse counter is reset by testInProgress when an operation is reused. During the third DataEntry window testInProgress had been temporarily disabled (SLM vs Karlsruhe difference), which allowed RefPulses to appear to persist.
2. Increase revision to 3.9.3134.113

Solution:
1. Restore/ensure testInProgress behavior during the DataEntry sequence and explicitly reset the pulse counter when starting a new operation.
2. Increased revision to 3.9.3134.113
2026-08-24 12:54:15 +02:00
marekf 7139cdbe80 Merge branch 'develop/SLM-PT50-7-2026-merged-smart-cam-slm-mex' into develop/SLM-PT50_genesisDirectDecode 2026-08-04 16:31:57 +02:00
marekf 93d7456ccd <Fix>: RefPulses carry-over by restoring testInProgress behavior in StandingStartMassCollection, increase revision to 3.9.3134.112
Cause:
1. Background: there was a suspicion that the Bukoven board left RefPulses active across operations; this was investigated and ruled out. The pulse counter is reset by testInProgress when an operation is reused. During the third DataEntry window testInProgress had been temporarily disabled (SLM vs Karlsruhe difference), which allowed RefPulses to appear to persist.
2. Increase revision to 3.9.3134.112

Solution:
1. Restore/ensure testInProgress behavior during the DataEntry sequence and explicitly reset the pulse counter when starting a new operation.
2. Increased revision to 3.9.3134.112
2026-08-04 16:21:33 +02:00
marekf 79de40515a Upgrade: Deleted LANG_PL compilation condition, disabled legalizator before test, and version to 3.9.3134.111
Cause:
1. Unnecessary and mandatory user verification before the test, in addition, a bug appears: the "Approval by legalizer" window is displayed behind the main application window. The window was only needed for the Polish project.
2. `LANG_PL` constant in `TBF.csproj` was no longer needed.
3. Assembly version and file version in `AssemblyInfo.cs` needed to be incremented to `3.9.3134.111`.

Solution:
1. Made `altProcNameLabel` and `altProcNameTextBox` always visible and disabled specific UI controls in `ProcedureDlg.cs`.
2. Removed `LANG_PL` from `DefineConstants` in `TBF.csproj`.
3. Updated assembly version and file version in `AssemblyInfo.cs` to `3.9.3134.111`.
2026-07-28 11:49:37 +02:00
marekf e38c6c1199 Upgrade: Add remaining time display for tests, version 3.9.3134.110
Cause:
1. Missing functionality to display remaining time during tests in `StandingStartMassCollectionSeq.cs` and `StandingStartMassCollectionAdvanceSeq.cs`.
2. Assembly version and file version required an update to reflect changes.

Solution:
1. Implemented remaining time display in `StandingStartMassCollectionSeq.cs` and `StandingStartMassCollectionAdvanceSeq.cs` using `Bridge.OnActivity` to show time in minutes and seconds or seconds only.
2. Updated assembly version and file version in `AssemblyInfo.cs` to `3.9.3134.110`.
2026-07-28 11:32:53 +02:00
marekf dbb35a203e Update: Adjust input unit constants in DataEntry form and increment version to 3.9.3134.109
Cause:
1. Updated `unit` constant in `TestStartEndForm` class within `TBF.Rig.DataEntry.StandardCamera` namespace from `Unit.m3` to `Unit.l`.
2. Updated `unit` constant in `TestStartEndForm` class within `TBF.Rig.DataEntry.StandartCameraPurchaseOrder` namespace from `Unit.m3` to `Unit.l`.
3. Incremented assembly version and file version in `AssemblyInfo.cs` to `3.9.3134.109`.

Solution:
1. Changed `unit` constant in `TestStartEndForm` class within `TBF.Rig.DataEntry.StandardCamera` namespace to `Unit.l`.
2. Changed `unit` constant in `TestStartEndForm` class within `TBF.Rig.DataEntry.StandartCameraPurchaseOrder` namespace to `Unit.l`.
3. Updated assembly version and file version in `AssemblyInfo.cs` to `3.9.3134.109`.
2026-07-28 07:38:30 +02:00
marekf a88a094e86 Release SLM-PT50 3.9.3134.108, increase revision after merge 2026-07-23 13:08:52 +02:00
marekf ffaebbdf07 Merge branch 'develop/SLM-PT50_genesisDirectDecode' of http://87.197.120.129:33300/michal/tbf into develop/SLM-PT50 2026-07-23 12:57:20 +02:00
marekf a210a0d1f2 Upgrade (UniDataStorageWriter): improve writer configuration UX and CSV handling
- added technology-specific default configuration for new writers
- auto-fill default SQL/XLSX write templates
- disabled write templates for CSV technology
- added technology-specific data source hints
- changed test parameter format to Column=Value and Where=Value;Set=Value
- improved parameter validation to prevent invalid input crashes
- added automatic CSV header creation and header validation
- aligned CSV writer behaviour with XLSX writer
- Increase revision number
2026-07-21 11:18:11 +02:00
marekf 4cd191fcb2 Merge branch 'develop/SLM-PT50-7-2026-merged-smart-cam-slm-mex' into develop/Mexico-6-2026 2026-07-20 15:54:01 +02:00
marekf e21833fb5f Release Mexico 3.9.3069.1 2026-07-20 15:47:06 +02:00
marekf b746ceb34a Upgrade - Replace PulsesTypeByQuantity with MultiFunctionalVariables unit filtering based on RegisterReader quantity type - remove PulsesTypeByQuantity-based parameter values
- generate available units using Units.IsQuantity(...)
- use Quantity.MultiFunctionalVariables to populate supported units
- centralize unit selection through Units.GetQuantity()
2026-07-20 14:09:29 +02:00
marekf 83c1f0e817 Update - Actualized revision nr. 2026-07-17 14:15:22 +02:00
marekf a88a5a8db2 Fix - 1. StandingStartMassCollection and StandingStartMassCollectionAdvance - changed all delayed Activity strings, 2. Added UI/Bridge onDelayedActivity() method for universal using 2026-07-17 14:15:04 +02:00
marekf 271200a8da Upgrade - StandingStartMassCollectionAdvance method - 1. changed structure and residual flow delays by changes in StandingStartMassCollection method, 2. added variables, counting and evaluation for volume/mass type of RegisterReader 2026-07-17 12:42:07 +02:00
marekf b703297f5e Fix - 1. repaired StandingStartMassCollection MassEN evaluation, the problem was caused by the end scale measuring only after recalculating the basic values - 2. removed fixed delay for the first diverter switching and first diverter switching was moved after a configurable delay - 3. hidden unitComboBox in the StartEnd dialog of the DataEntry.Standard24 component, because units are now defined using the RegisterReader component 2026-07-17 09:55:03 +02:00
marekf 637d1d021d Fix - StandingMassCollection method - add end mass measurement 2026-07-16 15:19:28 +02:00
marekf dbe424b17a Add - StandingStart method - Store and log default PumpPower for dynamic test repetitions - Read PumpPower from the current pump before starting dynamic tests.
- Store the value as the default PumpPower for subsequent test instances in the cycle.
- Add logging when PumpPower is initialized from the pump.
- Add logging at test start including test name, repetition number,
  last repetition flag, and PumpPower to improve diagnostics.
2026-07-16 14:21:00 +02:00
marekf be0525a319 Upgrade - improved StandingStart method by SLM requirements, added fixed delays for meters stabilization 2026-07-16 13:29:42 +02:00
marekf 0ce130f0ef Fix - repaired merging compatibility problem of GCI return type GciConnectResult - IsLoggedOn and IsConnected, increased VisualStudio version from 17.8 to 18.5 2026-07-16 12:37:59 +02:00
marekf 6ac3ea87ef Upgrade - improved StandingStartMassCollection method by SLM requirements, added fixed delays, added residual flow delays 2026-07-16 09:30:59 +02:00
marekf 38973e0030 Add - Store and log default PumpPower for dynamic test repetitions - Read PumpPower from the current pump before starting dynamic tests.
- Store the value as the default PumpPower for subsequent test instances in the cycle.
- Add logging when PumpPower is initialized from the pump.
- Add logging at test start including test name, repetition number,
  last repetition flag, and PumpPower to improve diagnostics.
2026-07-15 11:35:32 +02:00
marekf 37c89a4c9c Fix - missing safe shutdown when PerformSteps returns UiCmdStop
- When STOP is pressed during PerformSteps() in the BeforeTest transition,
Transition() returned immediately and skipped the common shutdown logic.
This caused FM pumps and valves to remain in their previous state.
- Perform the required safe shutdown (FM pump off and default valve setup)
before returning UiCmdStop or Error.
2026-07-14 14:41:53 +02:00
marekf 0075362f92 Fix - Handle scale underload condition
- Added Underload measurement state
- Detect underload responses from Mettler Toledo scales
- Automatically zero the scale after underload detection
- Restart mass measurement after successful zeroing
- Prevent zeroing response from being used as a valid measurement
- Preserve measurement stability by clearing buffered readings after underload
2026-07-13 14:01:57 +02:00
marekf 38d28599ad Merge branch 'develop/SLM-PT50_genesisDirectDecode' of http://87.197.120.129:33300/michal/tbf into develop/SLM-PT50 2026-07-03 13:37:26 +02:00
marekf c9a4b0f3ce Add (develop) - component ResultsWriter - added diag + english comments 2026-07-03 13:35:39 +02:00
marekf b74f1fd644 Add (develop) - first succesfull test - new component ResultsWriter - note: IResultsProcessor type componet, store result variables to database by using of component UniDataStorageWriter to the user-configured storage. The selection of variables is done using the standard UI for selecting the display of measurement results . In other words: The measurement results selected by the user are saved to a user-defined storage. 2026-07-03 12:46:56 +02:00
marekf 110b880304 Add (release) new component Modbus.WaterAnalyzerUni derived from original component Modbus.WaterAnalyzer. New component is expanded by the possibility to choose conductivity reading directly from the sensor (Modbus protocol) (original solution) or using a 4-20mA converter to serial line (RS232/485) (Modbus protocol). 2026-07-02 09:01:00 +02:00
marekf 7867044c0a Add (develop) new component ConductivityMeter, because we have papouch AD4RS converter (4-20mA to modbus) - first correct version for testing 2026-07-01 15:58:36 +02:00
marekf 88903f3dac Merge branch 'feature/task/Cordonel-preadjustmentUI-GenesisMeter-ZeroFlowGenesisMeter' into develop/UnionTown 2026-06-12 10:44:51 +02:00
marekf 4e223549ee Develop - GciBridge -> GCI -> PreadjustmentUI functions 7 - successful processes - testbench is exported to UnionTown - without CalibParams DB writer 2026-06-12 10:43:34 +02:00
marekf 2ea0df8faf Develop - GciBridge -> GCI -> create database core in GCI - SkeletoKey - CalibParams - first ok 2026-06-09 19:55:36 +02:00
marekf 061e85ca6b Develop - GciBridge -> GCI -> PreadjustmentUI functions 6 - first successful processes - without CalibParams DB reader and CalibParams DB writer 2026-06-09 08:53:54 +02:00
marekf 269e650b38 Merge branch 'feature/task/SWDocumentation-DocFX' into develop/UnionTown 2026-06-08 10:29:04 +02:00
marekf a3bfc7a43e Develop - GciBridge -> GCI -> PreadjustmentUI functions 5 2026-06-05 15:21:18 +02:00
marekf c2ff40d816 Develop - GCI - create documentation model 2, using DocFX tool - using .md files - output as html page 2026-06-02 07:26:36 +02:00
marekf eb09ae2eca Develop - GCI - create documentation model, using DocFX tool - using .md files - output as html page 2026-05-29 09:49:28 +02:00
marekf d1b6247bb7 Develop - GciBridge -> GCI -> create database core in GCI - SkeletoKey - CalibParams 2026-05-29 09:43:51 +02:00
marekf c5f52e06fa Develop - GciBridge -> GCI -> PreadjustmentUI functions 4 2026-05-26 10:51:18 +02:00
marekf 9700ba7900 Develop - GciBridge -> GCI -> PreadjustmentUI functions 3 2026-05-22 08:52:40 +02:00
marekf bc72130d09 Develop - GciBridge -> GCI -> PreadjustmentUI functions 2 2026-05-21 08:16:02 +02:00
marekf ad208f9633 Develop - GciBridge -> GCI -> PreadjustmentUI functions 2026-05-18 19:20:10 +02:00
marekf ad2bfd3639 Upgrade (develop) - GCI - GciBridge - improved Disconnect 2026-05-14 13:05:21 +02:00
marekf 5e637ae02e Upgrade (develop) - GCI - GciBridge - full parallel threading slots actions 2026-05-13 09:30:27 +02:00
marekf 63f8dd7467 Upgrade (develop) - GCI - GciBridge - InitSlot() - only ports config update if slot already exists 2026-05-12 14:59:15 +02:00
marekf aac3af8714 Upgrade (develop) - GCI - GciBridge - delete meter.Logout() in GetPcbId() 2026-05-12 14:20:24 +02:00
marekf 378917f59d Upgrade (develop) - GCI - GciBridge - GCI-Laatzen interface debugging 2026-05-12 13:30:21 +02:00
marekf 76e7ccb4a6 Update (develop) - GCI - GciBridge - better return for Connect 2026-05-11 17:14:56 +02:00
marekf 953b02a6e8 Upgrade (develop) - GCI - GciBridge - new CleanSlot() - Disconnect Hard Dispose 2026-05-08 11:49:14 +02:00
marekf 71a66dca2d File modified due to the need for MF to compile in advance, but the change is ultimately not necessary for either MF or MB 2026-05-07 18:23:50 +02:00
marekf 978c620287 Update (develop) - GCI - GciBridge - improved PublicModels - fixed bugs 2026-05-07 18:20:18 +02:00
marekf 8489dafdf7 Update (develop) - GCI - GciBridge - progress 2026-05-06 14:58:44 +02:00
marekf c1573390b5 Merge branch 'develop/UnionTown' into develop/SLM-PT50 2026-05-06 09:08:09 +02:00
marekf 5d60523e87 Fixed - next issue with old Newtonsoft.Json v 12.0 - updated to 13.0 2026-05-06 07:38:49 +02:00
marekf 91c63e0792 Fixed - bad copying GenesisCordonelInterface runtime files to TBF target folder 2026-05-06 07:35:17 +02:00
marekf 0940df2f1a Fixed - issue with old Newtonsoft.Json v 12.0 - updated to 13.0 2026-05-06 07:32:55 +02:00
marekf 7101fd679e Merge branch 'develop/UnionTown' into develop/SLM-PT50 2026-05-04 15:45:07 +02:00
marekf 86299da682 Fixed - bad runtime files copying in postbuilding of GenesisCordonelInterface project. Copying is replaced from TBF/Build/ to GCI/Build 2026-05-04 15:40:05 +02:00
marekf 39b951b709 Merge branch 'develop/UnionTown' into develop/SLM-PT50 2026-05-04 13:08:10 +02:00
marekf a475bd80f3 Change - ExternalProjects folder to gitignore 2026-05-04 13:07:28 +02:00
marekf 4e64bcd117 Merged branch Uniontown - GenesisCordonelInterface + GciBridge 2026-05-04 12:49:02 +02:00
marekf fbd0219a4a Update (develop) GCI - GciBridge - big change - for implement to GenesisHead and SmartReader 2026-05-04 11:29:13 +02:00
marekf 7e95e18207 Fixed - genesis library path issue 2026-04-27 16:52:50 +02:00
marekf 38fa86d243 Merge branch 'develop/UnionTown' into develop/SLM-PT50 2026-04-27 16:42:47 +02:00
marekf 4c1ce73de9 Upgrade (develop) - GCI - GciBridge - big change 2026-04-27 16:40:42 +02:00
marekf 15dc325275 Upgrade (develop) - GCI - big change 2026-04-27 07:57:16 +02:00