iPerl.WaterMeter: Q2 correction with limits given by Q2adj error limits
This commit is contained in:
@@ -110,7 +110,7 @@ namespace TBF.BenchControl.WaterMeters.iPerl
|
||||
if (volumeMeter > 1E-2)
|
||||
{
|
||||
UInt16 newFactor = (UInt16)((double)originalCalibrationFactor * VolumeLtrRef / volumeMeter + 0.5);
|
||||
log.InfoFormat("CalculateNewCalibFactor: orig={0} new={1} V_1={2} V_2={3} Vdiff={4} Vref={5}",
|
||||
log.InfoFormat("Calibration factor: orig={0} new={1} V_1={2} V_2={3} Vdiff={4} Vref={5}",
|
||||
originalCalibrationFactor, newFactor,
|
||||
VolumeLtrStart.ToString("F3"), VolumeLtrEnd.ToString("F3"), volumeMeter.ToString("F3"),
|
||||
VolumeLtrRef.ToString("F3"));
|
||||
@@ -122,7 +122,7 @@ namespace TBF.BenchControl.WaterMeters.iPerl
|
||||
}
|
||||
else
|
||||
{
|
||||
log.ErrorFormat("CalculateNewCalibFactor: orig={0} new={0} (unchanged!) Vdiff={1}",
|
||||
log.ErrorFormat("Calibration factor: orig={0} new={0} (unchanged!) Vdiff={1}",
|
||||
originalCalibrationFactor, volumeMeter.ToString("F3"));
|
||||
return originalCalibrationFactor; /// Too small volume in the denominator -> no correction at all
|
||||
}
|
||||
@@ -145,7 +145,24 @@ namespace TBF.BenchControl.WaterMeters.iPerl
|
||||
double F = D / (NominalTestFlow * 10.0); /// Error corrected with 8 Raw Units per minute [%]
|
||||
double G = F / B; /// Error corrected with 1 Raw Unit per minute [%]
|
||||
|
||||
return (-1) * (LastTestResult.VolumeErrorPct / G) * (LastTestResult.VolumeRef / LastTestResult.VolumeMeter);
|
||||
float errLimitLo = LastTestResult.TestResult.ErrLimLo;
|
||||
float errLimitHi = LastTestResult.TestResult.ErrLimHi;
|
||||
|
||||
float volumeMeterErrLimLo = LastTestResult.VolumeRef * (100.0f + errLimitLo) / 100.0f;
|
||||
float volumeMeterErrLimHi = LastTestResult.VolumeRef * (100.0f + errLimitHi) / 100.0f;
|
||||
|
||||
double corrFactorHi = (-1) * (errLimitLo / G) * (LastTestResult.VolumeRef / volumeMeterErrLimLo); /// > 0
|
||||
double corrFactorLo = (-1) * (errLimitHi / G) * (LastTestResult.VolumeRef / volumeMeterErrLimHi); /// < 0
|
||||
|
||||
double corrFactor = (-1) * (LastTestResult.VolumeErrorPct / G) * (LastTestResult.VolumeRef / LastTestResult.VolumeMeter);
|
||||
double origCalulatedCorrFactor = corrFactor;
|
||||
|
||||
if (corrFactor < corrFactorLo) corrFactor = corrFactorLo;
|
||||
if (corrFactor > corrFactorHi) corrFactor = corrFactorHi;
|
||||
|
||||
log.WarnFormat("Q2 correction: WM={0}, LoLim={1}, HiLim={2}, calculated={3}, corrFactor={4}",
|
||||
Name, corrFactorLo, corrFactorHi, origCalulatedCorrFactor, corrFactor);
|
||||
return corrFactor;
|
||||
}
|
||||
|
||||
/// <summary> Name set by the test, to be used as a part og the opto-data log file name </summary>
|
||||
|
||||
Reference in New Issue
Block a user