Cause:
- Preadjustment Completion and TBF Q3 calibration results required configurable database persistence.
- Both write operations required diagnostics and simulation support.
Solution:
1. Added independent database writer configurations
- Added PreadjustmentCompletionWriter and Q3CalibrationWriter to gci_config.
- Support SQL commands, stored procedures and parameter mappings.
- Added SQL table creation and verification scripts.
2. Added Preadjustment Completion storage
- Store PCB ID, meter size, amplitude settings, temperature offsets, zero-flow offsets, timestamp and overall result.
- Use captured Completion readout values.
- Preserve the original HTTP reporting.
- Report SQL connection and write status in the corresponding slot diagnostics.
3. Added Q3 calibration storage
- Forward TBF results through GCI Bridge to the GCI database API.
- Store signed deviations and written calibration factors for all three paths.
- Write to SQL after successful calibration register writes and StoreCalibration.
- Include PCB ID and UTC timestamp.
4. Added simulation support
- Generate sample Completion and Q3 records for simulated slots.
- Perform real SQL writes when the corresponding writer is enabled.
- Log connection progress, write outcomes and generated record IDs.
5. Separated calibration and database outcomes
- Preserve meter operation results when database storage fails.
- Expose database write status independently.
6. Increased revision
- Updated revision from 3.9.3146.108 to 3.9.3146.109.
Validation:
- Confirmed successful SQL storage for simulated Completion.
- Confirmed successful Q3 calibration result storage.
ToDo:
- Use the GCI Q3 storage API as the foundation for future migration of Q3 calibration from the direct TBF communication branch to GCI.