marekf a0d53f9ec9 <Fix>: RefPulses carry-over by restoring testInProgress behavior in StandingStart, increase revision to 3.9.3145
Cause:
1. Background: Restore/ensure testInProgress behavior before the DataEntry sequence and explicitly prevent to reset the pulse.
2. Increase revision to 3.9.3145

Solution:
1. Deleted .AddOperation(testInProgress) calling
2. Increased revision to 3.9.3145
2026-08-25 13:49:49 +02:00
marekf 1e45f48c75 <Update>: Version to 3.9.3144.0 in AssemblyInfo files 2026-08-25 08:23:27 +02:00
michal 26db172153 Camera WM Purchase Order - crop disabled 2026-08-24 12:21:49 +02:00
michal 2e75f65b4a Update version to 3.9.3142.0 in AssemblyInfo files 2026-08-20 00:27:39 +02:00
michal b904d0ca41 Fix Flip Mode - ver. 3.9.3141.0
Add robust handling for form closure in UNI Data Entry: prevent operations on disposed/closing forms, enhance logging, and update logic in serial number and volume updates. Increment version to 3.9.3141.0.
2026-08-19 19:13:41 +02:00
michal f34046df2a Update version to 3.9.3140.0 in AssemblyInfo files - FIX DataEntry.UNI 2026-08-19 14:49:24 +02:00
michal a4ef208233 Fix ver. 3.9.3139 iPerlCommunicationForm.cs problems + exceptions if another regreadrer has mistakely same ports
Enhance iPerl dialog layout and error handling: improve `ShuffleTextBoxes` with positional mapping, refine checkbox state tracking, add detailed logging for labels and checkboxes, and handle serial port exceptions. Increment version to 3.9.3139.0.
2026-08-18 16:09:21 +02:00
michal f3e5527145 ver. 3.9.3138.0 - OPTHO FORMATING improve (ASIC,genesis) 2026-08-17 10:27:10 +02:00
michal 3488939995 FIX iPerl ASIC meter - mexico
Update MultiplyFactor constant to 1 and increment version to 3.9.3137.0
2026-08-14 14:14:33 +02:00
michal 0f2ef5a5bf bugfix IPERL_ASIC conductivity
update mappings , version 3.9.3136.0
2026-08-13 15:16:58 +02:00
michal 55d2381e8a Mexico - read additional parameters and store to DB, ver. 3.9.3135.0
Add support for reading and handling additional common parameters in OptoHeadTest. Extend RadioService and related classes with enhanced parsing and validation logic. Update UI and constants for functionality integration.
2026-08-12 16:20:16 +02:00
marekf 93d7456ccd <Fix>: RefPulses carry-over by restoring testInProgress behavior in StandingStartMassCollection, increase revision to 3.9.3134.112
Cause:
1. Background: there was a suspicion that the Bukoven board left RefPulses active across operations; this was investigated and ruled out. The pulse counter is reset by testInProgress when an operation is reused. During the third DataEntry window testInProgress had been temporarily disabled (SLM vs Karlsruhe difference), which allowed RefPulses to appear to persist.
2. Increase revision to 3.9.3134.112

Solution:
1. Restore/ensure testInProgress behavior during the DataEntry sequence and explicitly reset the pulse counter when starting a new operation.
2. Increased revision to 3.9.3134.112
2026-08-04 16:21:33 +02:00
marekf 79de40515a Upgrade: Deleted LANG_PL compilation condition, disabled legalizator before test, and version to 3.9.3134.111
Cause:
1. Unnecessary and mandatory user verification before the test, in addition, a bug appears: the "Approval by legalizer" window is displayed behind the main application window. The window was only needed for the Polish project.
2. `LANG_PL` constant in `TBF.csproj` was no longer needed.
3. Assembly version and file version in `AssemblyInfo.cs` needed to be incremented to `3.9.3134.111`.

Solution:
1. Made `altProcNameLabel` and `altProcNameTextBox` always visible and disabled specific UI controls in `ProcedureDlg.cs`.
2. Removed `LANG_PL` from `DefineConstants` in `TBF.csproj`.
3. Updated assembly version and file version in `AssemblyInfo.cs` to `3.9.3134.111`.
2026-07-28 11:49:37 +02:00
marekf e38c6c1199 Upgrade: Add remaining time display for tests, version 3.9.3134.110
Cause:
1. Missing functionality to display remaining time during tests in `StandingStartMassCollectionSeq.cs` and `StandingStartMassCollectionAdvanceSeq.cs`.
2. Assembly version and file version required an update to reflect changes.

Solution:
1. Implemented remaining time display in `StandingStartMassCollectionSeq.cs` and `StandingStartMassCollectionAdvanceSeq.cs` using `Bridge.OnActivity` to show time in minutes and seconds or seconds only.
2. Updated assembly version and file version in `AssemblyInfo.cs` to `3.9.3134.110`.
2026-07-28 11:32:53 +02:00
marekf dbb35a203e Update: Adjust input unit constants in DataEntry form and increment version to 3.9.3134.109
Cause:
1. Updated `unit` constant in `TestStartEndForm` class within `TBF.Rig.DataEntry.StandardCamera` namespace from `Unit.m3` to `Unit.l`.
2. Updated `unit` constant in `TestStartEndForm` class within `TBF.Rig.DataEntry.StandartCameraPurchaseOrder` namespace from `Unit.m3` to `Unit.l`.
3. Incremented assembly version and file version in `AssemblyInfo.cs` to `3.9.3134.109`.

Solution:
1. Changed `unit` constant in `TestStartEndForm` class within `TBF.Rig.DataEntry.StandardCamera` namespace to `Unit.l`.
2. Changed `unit` constant in `TestStartEndForm` class within `TBF.Rig.DataEntry.StandartCameraPurchaseOrder` namespace to `Unit.l`.
3. Updated assembly version and file version in `AssemblyInfo.cs` to `3.9.3134.109`.
2026-07-28 07:38:30 +02:00
marekf a88a094e86 Release SLM-PT50 3.9.3134.108, increase revision after merge 2026-07-23 13:08:52 +02:00
marekf a210a0d1f2 Upgrade (UniDataStorageWriter): improve writer configuration UX and CSV handling
- added technology-specific default configuration for new writers
- auto-fill default SQL/XLSX write templates
- disabled write templates for CSV technology
- added technology-specific data source hints
- changed test parameter format to Column=Value and Where=Value;Set=Value
- improved parameter validation to prevent invalid input crashes
- added automatic CSV header creation and header validation
- aligned CSV writer behaviour with XLSX writer
- Increase revision number
2026-07-21 11:18:11 +02:00
marekf b746ceb34a Upgrade - Replace PulsesTypeByQuantity with MultiFunctionalVariables unit filtering based on RegisterReader quantity type - remove PulsesTypeByQuantity-based parameter values
- generate available units using Units.IsQuantity(...)
- use Quantity.MultiFunctionalVariables to populate supported units
- centralize unit selection through Units.GetQuantity()
2026-07-20 14:09:29 +02:00
marekf 83c1f0e817 Update - Actualized revision nr. 2026-07-17 14:15:22 +02:00
marekf 271200a8da Upgrade - StandingStartMassCollectionAdvance method - 1. changed structure and residual flow delays by changes in StandingStartMassCollection method, 2. added variables, counting and evaluation for volume/mass type of RegisterReader 2026-07-17 12:42:07 +02:00
marekf 637d1d021d Fix - StandingMassCollection method - add end mass measurement 2026-07-16 15:19:28 +02:00
marekf dbe424b17a Add - StandingStart method - Store and log default PumpPower for dynamic test repetitions - Read PumpPower from the current pump before starting dynamic tests.
- Store the value as the default PumpPower for subsequent test instances in the cycle.
- Add logging when PumpPower is initialized from the pump.
- Add logging at test start including test name, repetition number,
  last repetition flag, and PumpPower to improve diagnostics.
2026-07-16 14:21:00 +02:00
marekf be0525a319 Upgrade - improved StandingStart method by SLM requirements, added fixed delays for meters stabilization 2026-07-16 13:29:42 +02:00
michal 28607e2729 fix Strings - Camera CJMS11 - support dialog multi pictures 2026-07-16 12:28:28 +02:00
marekf 0075362f92 Fix - Handle scale underload condition
- Added Underload measurement state
- Detect underload responses from Mettler Toledo scales
- Automatically zero the scale after underload detection
- Restart mass measurement after successful zeroing
- Prevent zeroing response from being used as a valid measurement
- Preserve measurement stability by clearing buffered readings after underload
2026-07-13 14:01:57 +02:00
michal 26b7aa7e8b correction atempt MultiplyFactor for timestamp, volume calculation, 2026-07-02 23:53:15 +02:00
michal 92f63e744d fix INIT - Overload time fix for OptoTelegramRawTest 2026-07-02 23:24:00 +02:00
michal b0581d81ef Extend measurement window for low-flow tests 2026-07-02 23:08:28 +02:00
michal 540b9af871 Overload time fix for OptoTelegramRawTest 2026-07-02 22:32:19 +02:00
michal d11f7107fe fix time calculation from raw data 2026-07-02 21:49:01 +02:00
michal 3c5a28f69d changed LED state from 4 to 7 in 2026-07-02 21:47:14 +02:00
michal 16a76720b8 Update references and paths for Xylem.Common components, refine OptoHeadTest handling for connection requests, and increment assembly version to 3.9.3063.1. 2026-05-05 08:34:21 +02:00
michal 33a5e479a9 Refactor RadioService to improve asynchronous handling, replace legacy code with CommInterfaceBridge, and standardize methods for GenesisHead communication. Remove unused GenesisCommunication components and increment assembly version to 3.9.3062.1. 2026-04-28 16:26:26 +02:00
marekf 38fa86d243 Merge branch 'develop/UnionTown' into develop/SLM-PT50 2026-04-27 16:42:47 +02:00
marekf 6199bd48c5 Upgrade .NET Framework 4.7.2 to 4.8 for Laatzen + GCI + TBF compatibility 2026-04-24 13:00:53 +02:00
michal de6dd45dcb Fix Flow - Increment assembly version to 3.9.3059.1. 2026-04-23 09:56:32 +02:00
michal 842f009b42 Increment assembly version to 3.9.3058.1 and fix logging format issue in FlowMeter initialization. 2026-04-21 07:43:35 +02:00
michal 5a8478e782 Refactor GenesisSmartReader Q3 calibration tests and logic:
- Expand Q3 calibration tests with diverse scenarios in `GenesisSmartReader_Q3Test`.
- Introduce channel-specific validation for calibration tests.
- Adjust methods for Q3 calibration in `GenesisSmartReader` to support per-channel factors.
- Add utilities for preparing simulation data and refinements for test consistency.
- Update `CliRunner` with improved task management, timeout handling, and logging enhancements.
2026-04-17 16:05:16 +02:00
michal 33ef914984 Update GenesisReaderTests: add real input parsing, format variation handling, and rollover cases; remove outdated Q3 calibration tests; increment version to 3.9.3056.1 2026-04-15 09:56:54 +02:00
michal 93459ccfab Refactor water meter handling in SequenceBase and FlyingStartMassCollectionSeq:
- Enhance logic for enabling/disabling channels based on water meter data.
- Introduce Q3 calibration validation in `GenesisSmartReader` with configurable thresholds.
- Update `GenesisCfgCtrl` UI to disable certain inputs.
2026-04-14 15:41:13 +02:00
michal 009aefb7c9 Refactor GenesisSmartReader and SequenceBase logic, enhance channel handling in water meters, and update GUI components:
- Adjust `Disabled` logic for water meters in `SequenceBase` allowing conditional enabling of channels based on serial number.
- Introduce `EnableShowChanels` property in `GenesisSmartReader` and integrate into `FlyingStartMassCollectionSeq`.
- Modify `StartupFlushMs` to derive configuration dynamically.
- Update `GenesisCfgCtrl` with a checkbox for enabling channels and input for flush timing.
- Increment assembly version to `3.9.3054.1`.
2026-04-14 08:03:02 +02:00
michal d2e18d615d Bump assembly version to 3.9.3053.1. 2026-04-13 14:17:56 +02:00
michal 2210bb8240 Enhance FlyingStartMassCollectionSeq with GenesisSmartReader support; add water meter handling logic and bump assembly version to 3.9.3050.1. 2026-04-11 21:55:34 +02:00
michal 2136695b80 Fix Stop 2026-04-08 16:37:27 +02:00
michal c0f9b0ffa0 Refactor GenesisSmartReader stop logic for enhanced thread safety and post-processing; bump assembly version to 3.9.3047.1. 2026-04-08 15:44:59 +02:00
michal c4ef22436f Update GenesisSmartReader Stop logic and bump assembly version to 3.9.3045.1:
- Refactor `Stop` for improved thread safety using `lock` on `_stateSync` and `_stopSync`.
- Enhance `Stop` behavior by introducing state updates during immediate stop requests.
- Fix redundant operations and reorganize log and processing statements for clarity.
- Update `RequestImmediateStopInternal` to accept a new state parameter.
2026-04-08 15:06:45 +02:00
michal 0b73871518 Refactor GenesisSmartReader and add advanced test utilities:
- Introduce `ForceProcessingEnabled`, `InstallSuccessfulProcessingOverride`, and `EnqueueIncomingLine` helpers to streamline test setup.
- Add new test suite `GenesisSmartReaderStopTests` to validate `Stop` behavior for different processing states.
- Enhance `GenesisSmartReader` state management with a dedicated `Reset` method and improved `Stop` logic.
- Improve buffer initialization in `ProcessOptoLine` to address null-value edge cases.
- Update assembly version to `3.9.3043.1`.
2026-04-08 13:51:55 +02:00
michal dcb0bf9a88 Update GenesisSmartReader logging and bump assembly version to 3.9.3040.1:
- Enhance `Stop` logs with `registerReader` and `optoSerialPort` details for improved traceability.
2026-04-07 15:27:20 +02:00
michal f6433134c1 Add WaitForStartupFlushToFinish and EnableProcessingLoopForTests utilities for improved test reliability:
- Incorporate `WaitForStartupFlushToFinish` in `GenesisSmartReaderThreadedReadTests` to ensure startup flush completion before processing.
- Introduce `EnableProcessingLoopForTests` in `GenesisSmartReaderTest` to streamline test initialization.
- Fix thread-safety issues in startup flush handling and reorder operations in `GenesisSmartReader.Stop`.
- Bump assembly version to `3.9.3039.1`.
2026-04-07 14:45:37 +02:00
michal 7e3df77b5b Refactor GenesisSmartReader processing logic and tests for robust handling:
- Replace property getters with concise expressions in `GenesisSmartReader`.
- Implement FIFO and concurrent producer-consumer unit tests for `ProcessingLoop`.
- Add startup flush mechanism for improved input handling on initialization.
- Refactor processing loop with structured start/stop tasks and cancellation tokens.
- Fix thread-safety issues and enhance logic in `Stop` and `StopDataStreamProcessing`.
- Add internal visibility for tests in `TBFTests` via `InternalsVisibleTo`.
- Update assembly version to `3.9.3038.1`.
2026-04-07 10:42:27 +02:00