(1) Dummy.Flowmeter displays flow measured by a scale, (2) OoR values are orange.

This commit is contained in:
Milan Hanajik
2021-04-26 11:56:01 +02:00
parent 870b49d8ad
commit fdb1e112dc
8 changed files with 161 additions and 38 deletions
+15 -7
View File
@@ -1,15 +1,16 @@
///
/// Copyright (c) 2017 Sensus Metering Systems
/// Copyright (c) 2017-2021 Sensus Slovensko a.s.
///
using System;
using System.Collections.Generic;
using log4net;
using SchematicDrawing;
using TBF.Boxes;
using TBF.Rig.Sequences;
namespace TBF.Rig.Dummy.FlowMeter
{
public class FlowMeter : ComponentBase, GenericDevices.IFlowMeter, IDrawingItCmpnt
public class FlowMeter : ComponentBase, GenericDevices.IFlowMeter, IDrawingItCmpntWithMeasuredVal
{
private static readonly ILog log = LogManager.GetLogger(typeof(FlowMeter));
public override string ToString() { return string.Format("{0}({1})", ClassName, Cfg.ToString(-1)); }
@@ -27,9 +28,16 @@ namespace TBF.Rig.Dummy.FlowMeter
public int Idx1 { get { return 1; } }
public bool MsrmntAvailable { get { return true; } }
public double MeasuredVal { get { return 1.234; } }
public string AltString { get { return "dummy"; } }
public bool MsrmntAvailable { get { return ProcessData.Devices != null && ProcessData.Devices.FlowMeter == this && ReadFlow() > 0; } }
public string AltString { get { return string.Empty; } }
public double MeasuredVal
{
get { return (ProcessData.Devices != null && ProcessData.Devices.FlowMeter == this) ? ReadFlow() : 0; }
set { }
}
public double MsrdValLimLo { get { return flowMeterCfg.MsrdValLimLo; } set { flowMeterCfg.MsrdValLimLo = value; } }
public double MsrdValLimHi { get { return flowMeterCfg.MsrdValLimHi; } set { flowMeterCfg.MsrdValLimHi = value; } }
public FlowMeter() { }
@@ -44,8 +52,8 @@ namespace TBF.Rig.Dummy.FlowMeter
public override void Initialize() { }
public double ReadFlow() { return 1.23; }
public double ReadFrequency() { return 1.23 * 2000.0 / (double)NominalFlow; }
public double ReadFlow() { return ProcessData.FlowFromMassIncrease; }
public double ReadFrequency() { return ProcessData.FlowFromMassIncrease * 2000.0 / (double)NominalFlow; }
/// <summary>
/// Events: FlowDone, Error
+17 -4
View File
@@ -1,5 +1,5 @@
///
/// Copyright (c) 2017 Sensus Metering Systems
/// Copyright (c) 2017-2021 Sensus Slovensko a.s.
///
using System.Collections.Generic;
using System.Xml.Serialization;
@@ -8,7 +8,7 @@ using TBF.Rig.Generic;
namespace TBF.Rig.Dummy.FlowMeter
{
public class FlowMeterCfg : ComponentCfgBase, IComponentCfg, IDrawingItem
public class FlowMeterCfg : ComponentCfgBase, IComponentCfg, IDrawingItemWithMeasuredVal
{
public static XmlSerializer Serializer = XmlSerializer.FromTypes(new[] { typeof(FlowMeterCfg) })[0];
public override XmlSerializer GetSerializer() { return Serializer; }
@@ -30,8 +30,21 @@ namespace TBF.Rig.Dummy.FlowMeter
public int LblY { get; set; }
public Orient LblOrient { get; set; }
public int MsrdX { get; set; }
public int MsrdY { get; set; }
public Orient MsrdOrient { get; set; }
public string MsrdFormat { get; set; }
public Config.Unit MsrdUnit { get; set; }
[XmlIgnore]
public IList<GNode> GNodes { get; set; }
[XmlIgnore]
public double MsrdValLimLo { get { return msrdValLimLo; } set { msrdValLimLo = value; } }
[XmlIgnore]
public double MsrdValLimHi { get { return msrdValLimHi; } set { msrdValLimHi = value; } }
double msrdValLimLo;
double msrdValLimHi;
/// Private parameterless constructor invoked by all other (public) constructors
FlowMeterCfg()
@@ -44,14 +57,14 @@ namespace TBF.Rig.Dummy.FlowMeter
{
Shape = Shape.FlowM;
Sz = Sz.M;
MsrdFormat = "{0:F4}";
MsrdUnit = Config.Unit.m3ph;
Name = name;
Factory = factory;
NominalFlow = 1.0f;
}
public string ToString(int i)
{
return string.Format("{0} {1}[{2},{3}] NominalFlow={4}", Name, Shape, X, Y, NominalFlow);
+4 -5
View File
@@ -1,5 +1,5 @@
///
/// Copyright (c) 2017 Sensus Metering Systems
/// Copyright (c) 2017-2021 Sensus Slovensko a.s.
///
using System;
using log4net;
@@ -41,7 +41,8 @@ namespace TBF.Rig.Dummy.FlowMeter
/// <summary>Start this operation</summary>
public void Start()
{
}
if (flowBox != null) flowBox.Val = flowMeter.ReadFlow();
}
/// <summary>Run this operation</summary>
/// <returns>
@@ -54,8 +55,6 @@ namespace TBF.Rig.Dummy.FlowMeter
}
/// <summary>Stop this operation</summary>
public void Stop()
{
}
public void Stop() { }
}
}
+14 -2
View File
@@ -29,6 +29,7 @@ namespace TBF.Rig.Sequences
///------------------------------------------------------------
protected static Rig.BenchPath benchPath;
protected static Rig.OutputPath outPath;
public static Rig.OutputPath Devices { get { return outPath; } }
public static readonly IList<IDrawingItCmpntWithMeasuredVal> ComponentsWithMeasuredVal;
@@ -215,8 +216,10 @@ namespace TBF.Rig.Sequences
public static DoubleBox RefFrequency = new DoubleBox() { Name = "RefFreq", Format = "F2" };
public static DoubleBox RefFlow = new DoubleBox() { Name = "RefFlow", Format = "F2" }; /// [m3/h]
public static DoubleBox Mass = new DoubleBox() { Name = "Mass", Format = "F3" }; /// [kg]
public static DoubleBox StartMass = new DoubleBox() { Name = "Start Mass", Format = "F3" }; /// [kg]
public static DoubleBox EndMass = new DoubleBox() { Name = "End Mass", Format = "F3" }; /// [kg]
public static double FlowFromMassIncrease;
public static DateTime TestStartTime;
public static DateTime TestEndTime;
@@ -240,6 +243,7 @@ namespace TBF.Rig.Sequences
RefPulses = 0;
RefFrequency.Clear();
RefFlow.Clear();
Mass.Clear();
StartMass.Clear();
EndMass.Clear();
}
@@ -296,6 +300,7 @@ namespace TBF.Rig.Sequences
public static Statistics PressDeltaStat = new Statistics(3, 4, true);
public static Statistics ConductStat = new Statistics(0, 4, true, new Plotter("Conductivity"));
public static Statistics RefFlowStat = new Statistics(5, 7, true, new Plotter("Flow"));
public static Statistics MassStat = new Statistics(0, 7, false, new Plotter("Mass"));
public static Statistics TempRefHiStat = new Statistics();
public static Statistics TempRefLoStat = new Statistics();
@@ -327,8 +332,9 @@ namespace TBF.Rig.Sequences
PressDeltaStat.Start(batchNr, testName, repetition, skippedSamplesCount);
ConductStat.Start (batchNr, testName, repetition, skippedSamplesCount);
RefFlowStat.Start (batchNr, testName, repetition, skippedSamplesCount);
TempRefHiStat.Start (batchNr, testName, repetition);
MassStat.Start (batchNr, testName, repetition, skippedSamplesCount);
TempRefHiStat.Start(batchNr, testName, repetition);
TempRefLoStat.Start (batchNr, testName, repetition);
Energy.Start (batchNr, testName, repetition);
VolumeForEnergy.Start(batchNr, testName, repetition);
@@ -353,6 +359,10 @@ namespace TBF.Rig.Sequences
ConductStat.Update(Conductivity);
RefFlowStat.Update(RefFlow);
double massIncreasePerSec;
MassStat.Update(Mass, out massIncreasePerSec);
FlowFromMassIncrease = 3600 * massIncreasePerSec / Config.Formulas.DistilledWaterDensityFromTemp(TempDiv.Val);
if (BatchRslts.Batch.HeatMeter)
{
TempRefHiStat.Update((TempRefHi1.Val + TempRefHi2.Val) / 2);
@@ -377,6 +387,8 @@ namespace TBF.Rig.Sequences
PressDeltaStat.Stop();
ConductStat.Stop();
RefFlowStat.Stop();
MassStat.Stop();
FlowFromMassIncrease = 0;
TempRefHiStat.Stop();
TempRefLoStat.Stop();
+78 -14
View File
@@ -13,7 +13,8 @@ namespace TBF.Rig.Sequences
bool recordingInProgress;
double[] fifo;
double[] sorted;
double first;
double[] timeStamps;
double first;
double last;
double min;
double max;
@@ -46,6 +47,7 @@ namespace TBF.Rig.Sequences
{
fifo = new double[FilterSize];
sorted = new double[FilterSize];
timeStamps = new double[FilterSize];
}
recordingInProgress = false;
@@ -117,6 +119,15 @@ namespace TBF.Rig.Sequences
Update(box.Val);
}
/// <summary>
/// Updates statistics
/// </summary>
/// <param name="value">New boxed value</param>
public void Update(TBF.Boxes.DoubleBox box, out double increasePerSecond)
{
Update(box.Val, out increasePerSecond);
}
/// <summary>
/// Updates statistics
/// </summary>
@@ -126,6 +137,15 @@ namespace TBF.Rig.Sequences
Update(box.Val);
}
/// <summary>
/// Updates statistics
/// </summary>
/// <param name="value">New boxed value</param>
public void Update(TBF.Boxes.FloatBox box, out double increasePerSecond)
{
Update(box.Val, out increasePerSecond);
}
/// <summary>
/// All samples are passed to this function
/// </summary>
@@ -136,22 +156,45 @@ namespace TBF.Rig.Sequences
{
if (totalCount >= skippedSamplesCount)
{
Process(value);
double dummy;
Process(value, out dummy, false);
}
totalCount++;
}
}
/// <summary>
/// All samples are passed to this function
/// </summary>
/// <param name="value">New value</param>
public void Update(double value, out double increasePerSecond)
{
increasePerSecond = 0;
if (recordingInProgress)
{
if (totalCount >= skippedSamplesCount)
{
Process(value, out increasePerSecond, true);
if (FilterSize > 1)
{
}
}
totalCount++;
}
}
/// <summary>
/// Only samples after first 'SkippedSamplesCount' samples are passed to this function.
/// Any filtering is done inside this function.
/// </summary>
/// <param name="value">New value</param>
void Process(double value)
void Process(double value, out double increasePerSecond, bool calculateIncrease = false)
{
double filteredValue;
if (ApplyFilter(value, out filteredValue))
if (ApplyFilter(value, out filteredValue, out increasePerSecond, calculateIncrease))
{
sum += filteredValue;
if (count == 0) first = filteredValue;
@@ -172,31 +215,43 @@ namespace TBF.Rig.Sequences
/// <param name="value">Input value</param>
/// <param name="filteredValue">Output filtered value</param>
/// <returns>true when 'fileterdValue' is valid</returns>
bool ApplyFilter(double value, out double filteredValue)
bool ApplyFilter(double value, out double filteredValue, out double increasePerSecond, bool calculateIncrease = false)
{
increasePerSecond = 0;
if (FilterSize == 0)
{
/// No filter => Pass the input value to the output
filteredValue = value;
return true;
}
}
/// Shift values in FIFO buffer, calculate the sum
double sum1 = 0;
for (int i = FilterSize - 1; i > 0; i--)
double sumX = 0;
double sumY = 0;
double sumXX = 0;
double sumXY = 0;
for (int i = FilterSize - 1; i > 0; i--)
{
fifo[i] = fifo[i - 1];
sum1 += fifo[i];
timeStamps[i] = timeStamps[i - 1];
///
sumX += timeStamps[i];
sumY += fifo[i];
sumXX += timeStamps[i] * timeStamps[i];
sumXY += timeStamps[i] * fifo[i];
}
/// Insert the current value into FIFO, add it to the sum
fifo[0] = value;
sum1 += value;
timeStamps[0] = Convert.ToDouble(StateMachine.Time);
///
sumX += timeStamps[0];
sumY += fifo[0];
sumXX += timeStamps[0] * timeStamps[0];
sumXY += timeStamps[0] * fifo[0];
fifoCount++;
if (fifoCount < FilterSize)
{
/// FIFO not filled yet => Pass the input value to the output
@@ -220,9 +275,18 @@ namespace TBF.Rig.Sequences
}
else
{
filteredValue = sum1 / FilterSize;
filteredValue = sumY / FilterSize;
}
if (calculateIncrease)
{
double D = FilterSize * sumXX - sumX * sumX;
if (D != 0)
{
increasePerSecond = (FilterSize * sumXY - sumX * sumY) / D;
}
}
return true;
}
}
@@ -1,5 +1,5 @@
///
/// Copyright (c) 2020 Sensus Slovensko a.s.
/// Copyright (c) 2020-2021 Sensus Slovensko a.s.
///
using System;
using System.Collections.Generic;
@@ -101,6 +101,12 @@ namespace TBF.Rig.TestMethods.FixedStartMassCollAdvanced
new List<Event> { Event.ConfigurationError };
}
}
else if (outPath.FlowMeter is TBF.Rig.Dummy.FlowMeter.FlowMeter)
{
var flowm = outPath.FlowMeter as TBF.Rig.Dummy.FlowMeter.FlowMeter;
flowm.MsrdValLimLo = test.Qfrom;
flowm.MsrdValLimHi = test.Qto;
}
IList<Event> e; /// Events from currently running operations
int tMass1 = 0;
@@ -1059,6 +1065,13 @@ namespace TBF.Rig.TestMethods.FixedStartMassCollAdvanced
while (!e.Contains(Event.PreviousStopped));
}
if (outPath.FlowMeter is TBF.Rig.Dummy.FlowMeter.FlowMeter)
{
var flowm = outPath.FlowMeter as TBF.Rig.Dummy.FlowMeter.FlowMeter;
flowm.MsrdValLimLo = 0;
flowm.MsrdValLimHi = flowm.NominalFlow;
}
return new List<Event> { retVal };
}
}