<Fix>: Correct GCI streaming port ownership for preadjustment, increase revision to 3.9.3146.104
Cause: - GenesisRegReader configured the GCI streaming port as NA. - GCI Bridge therefore received no calibration streaming data during the Amplitude Test. - The request port remained functional, so register writes succeeded while TOF data collection failed. Solution: 1. Corrected streaming port ownership - Pass the configured Genesis opto COM port to GCI Bridge as the UART streaming port. - Stop GenesisRegReader streaming before handing the port over to GCI Bridge. - Validate request and streaming ports during preadjustment detection. - Release the GCI connection after preadjustment so the port can be reused by subsequent tests. 2. Increased revision - Updated revision to 3.9.3146.104.
This commit is contained in:
@@ -1665,16 +1665,22 @@ namespace GenesisCordonelInterface.API
|
||||
|
||||
SetUnknownStatus(allMeterControls);
|
||||
|
||||
CheckNormalMeters(_meterControls, token);
|
||||
CheckTemperatureMeters(_progressProcess.Setting, _tempMeterControls, token);
|
||||
|
||||
CheckTemperatureMeters(
|
||||
_progressProcess.Setting,
|
||||
_tempMeterControls,
|
||||
token);
|
||||
|
||||
List<MeterStateControl> failedMeters = _meterControls.Where(m => m.IsEnabled && !m.Ok).ToList();
|
||||
|
||||
return new PreadjustmentDetectResult
|
||||
{
|
||||
Success = true,
|
||||
DetectedMeterCount = _meterControls.Count(m => m.IsEnabled),
|
||||
DetectedThermometerCount = _tempMeterControls.Count(t => t.IsEnabled)
|
||||
Success = failedMeters.Count == 0,
|
||||
DetectedMeterCount = _meterControls.Count(m => m.IsEnabled && m.Ok),
|
||||
DetectedThermometerCount = _tempMeterControls.Count(t => t.IsEnabled && t.Ok),
|
||||
ErrorMessage = failedMeters.Count == 0 ? null : $"Request or streaming port check failed for slots: {string.Join(",", failedMeters.Select(m => m.Slot))}."
|
||||
};
|
||||
}
|
||||
catch (Exception ex)
|
||||
|
||||
Reference in New Issue