VolumeForEnergy calculated in [l], ver. 2.10.341
This commit is contained in:
+3
-3
@@ -407,8 +407,8 @@ namespace TBF.BenchControl.TestMethods.FixedStartMassCollection
|
||||
//double deltaTime = (double)(StateMachine.Time - lastEnergyUpdateTime);
|
||||
float T_in = (heatMetersPath.TMeterRefWarm1.ReadTemperature() + heatMetersPath.TMeterRefWarm2.ReadTemperature()) / 2; /// [°C]
|
||||
float T_out = (heatMetersPath.TMeterRefCold1.ReadTemperature() + heatMetersPath.TMeterRefCold2.ReadTemperature()) / 2; /// [°C]
|
||||
double deltaVolume = 0.001 * LtrPerRefPulse * RefPulsesDelta; /// [m3]
|
||||
double deltaEnergy = deltaVolume * (T_in - T_out) * Formulas.HeatCoefficientWater(16, T_in, T_out, heatMetersTestParams.FlowMeasuredAtHiTempPipe); /// [J]
|
||||
double deltaVolume = LtrPerRefPulse * RefPulsesDelta; /// [l]
|
||||
double deltaEnergy = 0.001 * deltaVolume * (T_in - T_out) * Formulas.HeatCoefficientWater(16, T_in, T_out, heatMetersTestParams.FlowMeasuredAtHiTempPipe); /// [J] = [m3] * [K] * [J/(m3 K)]
|
||||
Energy.Update(deltaEnergy);
|
||||
VolumeForEnergy.Update(deltaVolume);
|
||||
lastEnergyUpdateTime = StateMachine.Time;
|
||||
@@ -677,7 +677,7 @@ namespace TBF.BenchControl.TestMethods.FixedStartMassCollection
|
||||
}
|
||||
else if (heatMetersTestParams != null)
|
||||
{
|
||||
tstRslt.RefEnergy = Energy.Sum * tstRslt.VolumeCTV * 0.001 / VolumeForEnergy.Sum;
|
||||
tstRslt.RefEnergy = Energy.Sum * tstRslt.VolumeCTV / VolumeForEnergy.Sum; /// [J] = [J] * [l] / [l]
|
||||
|
||||
///
|
||||
/// Heat meters
|
||||
|
||||
@@ -29,5 +29,5 @@ using System.Runtime.InteropServices;
|
||||
// Build Number
|
||||
// Revision
|
||||
//
|
||||
[assembly: AssemblyVersion("2.10.340.1")]
|
||||
[assembly: AssemblyFileVersion("2.10.340.1")]
|
||||
[assembly: AssemblyVersion("2.10.341.1")]
|
||||
[assembly: AssemblyFileVersion("2.10.341.1")]
|
||||
|
||||
Reference in New Issue