E1 error indicator modified, average flow calculated from CTV volume and test time verified, ver. 2.18.1019
This commit is contained in:
@@ -42,8 +42,8 @@ namespace TBF.BenchControl.Various.ErrorFlags
|
|||||||
/// Any E# is 'true' on error, 'false' when OK
|
/// Any E# is 'true' on error, 'false' when OK
|
||||||
|
|
||||||
/// Average flow
|
/// Average flow
|
||||||
bool E1 = ProcessData.RefFlowStat.Average < testRslt.Qfrom()
|
double aveFlow = 3.6 * testRslt.VolumeCTV / testRslt.TestTime;
|
||||||
|| ProcessData.RefFlowStat.Average > testRslt.Qto();
|
bool E1 = (aveFlow < testRslt.Qfrom()) || (aveFlow > testRslt.Qto());
|
||||||
|
|
||||||
/// Max. and min. up and down water temperature
|
/// Max. and min. up and down water temperature
|
||||||
bool E2 = ProcessData.TempUpStat.Min < testRslt.TempLimLo()
|
bool E2 = ProcessData.TempUpStat.Min < testRslt.TempLimLo()
|
||||||
|
|||||||
@@ -29,5 +29,5 @@ using System.Runtime.InteropServices;
|
|||||||
// Build Number
|
// Build Number
|
||||||
// Revision
|
// Revision
|
||||||
//
|
//
|
||||||
[assembly: AssemblyVersion("2.18.1018.0")]
|
[assembly: AssemblyVersion("2.18.1019.0")]
|
||||||
[assembly: AssemblyFileVersion("2.18.1018.0")]
|
[assembly: AssemblyFileVersion("2.18.1019.0")]
|
||||||
|
|||||||