Temperature (TempUpMean + TempDownMean) / 2 used for density and volume calculations, ver. 2.17.712

This commit is contained in:
Milan Hanajik
2017-10-19 19:04:13 +02:00
parent 0df833ca8a
commit 0c2839e6c9
4 changed files with 17 additions and 14 deletions
@@ -331,8 +331,9 @@ namespace TBF.BenchControl.TestMethods.ReferenceFlowmeterCalibration
tstRslt.FlowMass = 3600.0 * (tstRslt.MassEnd - tstRslt.MassStart) / tstRslt.TestTime; /// [kg/h]
tstRslt.FlowVolume = 3.6 * LtrPerRefPulse * tstRslt.PulsesMaster / tstRslt.TestTime; /// [m3/h]
tstRslt.Buoyancy = Formulas.Buoyancy();
tstRslt.VolumeCTV = 1000.0 * tstRslt.Buoyancy * (tstRslt.MassEnd - tstRslt.MassStart) / tstRslt.DensityOut; /// [l] 1000.0f is because density is in [kg/m3]
tstRslt.VolumeMaster = LtrPerRefPulse * tstRslt.PulsesMaster; /// [l] volume from the master flow meter
double density = Formulas.WaterDensityFromTemp((tstRslt.TempUpMean + tstRslt.TempDownMean) / 2.0);
tstRslt.VolumeCTV = 1000.0 * tstRslt.Buoyancy * (tstRslt.MassEnd - tstRslt.MassStart) / density; /// [l] 1000.0f is because density is in [kg/m3]
tstRslt.VolumeMaster = LtrPerRefPulse * tstRslt.PulsesMaster; /// [l] volume from the master flow meter
tstRslt.ConstMaster = LtrPerRefPulse * tstRslt.VolumeCTV / tstRslt.VolumeMaster; /// Corrected master pulses per liter
tstRslt.ErrorMaster = Formulas.ErrorFromVolumes(tstRslt.VolumeMaster, tstRslt.VolumeCTV);